Chapter 3. Operation of the Microscope
A1
11
3.6
Perfect Focus System Operation
(with the Front Operation Panel)
(1) Check that the dichroic mirror IN/OUT lever is
set to IN.
(2) Set the objective to be used in the optical path
and adjust the focus with the focus knob on
the microscope to focus on the sample.
(3) Press the PFS-ON button on the front
operation panel.
(When the PFS starts, the PFS-ON indicator
lights green.)
OUT
DICHROIC MIRROR
IN
Dichroic mirror IN/OUT lever
(4) Focus on the position at which focus is maintained with the offset dial on the PFS offset controller and
start microscopy.
Once the PFS operation is started, focusing with the focus knob on the microscope is disabled.
Registering offset amount: The focus position adjusted with the offset dial can be registered with the
PFS-MEMORY button.
Recalling offset amount:
Registered focus position can be called with the PFS-RECALL button.
L80
EYE
DISPLAY
MEMORY
PFS
ON
RECALL
Z-RESET
BRIGHTNESS
R100
L100
FOCUS
1X
1.5X
PFS
DOWN
UP
PFS
OFFSET
CONTROLLER
FINE
COARSE
Figure Ti-E Front Operation Panel and Offset Controller
PFS-ON
button/indicator
PFS-MEMORY
button/indicator
PFS-RECALL
button/indicator
FINE/COARSE selector
button