137
ADVANCED FEATURES
Parameter Settings
3
Standard accommodation power display (only for Plus Package): Morgan, Hofstetter, No
Factory setting: No
Setting for the standard accommodation power to be displayed in the # 19 Near point of accommoda-
tion (NPA) window and the #19 Accommodation with addition of minus lens window.
Morgan
Displays the range of Morgan’s criterion
Hofstetter
Displays the maximum value, average value, and minimum value calculated by the
Hofstetter formula
No
Criterion is not displayed.
Acomm. power for Range of Clear Vision: ADD/NPA/Age/Manual
Factory setting: Age
Setting for selecting which value is used for the accommodation power for the range of clear vision cal-
culation.
Near indication link (Range of Clear Vision): Yes, No
Factory setting: Yes
Setting for whether or not to switch the far / near indication automatically along with the lens type when
the lens type is changed with the data window displayed during the test for range of clear vision.
AC/A for Sheard/Percival SPH (only for Plus Package): Gradient AC/A, Heterophoria AC/A
Factory setting: Gradient AC/A
Setting for AC/A used for calculating the SPH adjustment value that satisfies Sheard/Percival’s crite-
rion.
Accomm. for Morgan’s System (only for Plus Package): #19 NPA, #19 (-)lens
Factory setting: #19 NPA
Setting for accommodation used for comparison with Morgan’s criterion.
Prism for Supra. and Infra. (only for Plus Package): R, L
Factory setting: R
Selection of which prism to be changed when in Supravergence test. For R, #12B and #18B result
becomes right supravergence / infravergence test. For L, the result becomes left supravergence /
infravergence test.
After measuring #12B or #18B, the setting cannot be changed.
Graph type (only for Plus Package): Type 1, Type 2
Factory setting: Type 1
Setting for the type of visual function analysis graph display.
Type 1
#20 data is not reflected on the RCI line. #21 data is not reflected to the RCS line.
Type 2
#20 data is reflected on the RCI line. #21 data is reflected to the RCS line.
*For details, see
"5.3.2 Visual function analysis graph" (page 237).
Graph display (only for Plus Package): As - +13.00D, As - +3.0D
Factory setting: As - +13.00D
Setting for the graph display.
161221687130 承認済
Uncontrolled copy when printed.
Summary of Contents for RT-5100
Page 12: ...X 161221687130 承認済 Uncontrolled copy when printed ...
Page 18: ... XVI 161221687130 承認済 Uncontrolled copy when printed ...
Page 46: ...28 BEFORE USE Getting Started and Exiting 161221687130 承認済 Uncontrolled copy when printed ...
Page 274: ...256 MAINTENANCE List of Replacement Parts 161221687130 承認済 Uncontrolled copy when printed ...
Page 302: ...284 EMC ELECTROMAGNETIC COMPATIBILITY 161221687130 承認済 Uncontrolled copy when printed ...