21 Point Exams (only for Plus Package)
21 Point Exams (only for Plus Package)
Standard accommodation power display
According to the “Standard accommodation power display” parameter, the maximum value, average
value, and minimum value calculated by the Morgan’s criterion or Hofstetter formula can be displayed
on the accommodation measurement window. To display this value, patient’s age must be entered
beforehand. For details, see "3.11 Age Input" (Page 118).
Specify the accommodation power to be displayed with the “Standard accommodation display”
parameter.
When the “Standard accommodation power display” parameter is set to
“Morgan”
Displays the range of Morgan’s criterion calculated by age.
Morgan’s criterion
accommodation power
*
When the accommodation power is below the range of criterion, or the value is 0(D), the accom-
modation power is displayed in pink.
When the “Standard accommodation power display” parameter is set to “Hof-
stetter”
Displays the minimum value, average value, and maximum value calculated by Hofstetter formula.
Minimum value,
average value,
maximum value
*
When the accommodation power (D) is below the minimum value, or the value is 0(D), the accom-
modation power is displayed in pink.
161221687130 承認済
Uncontrolled copy when printed.
Summary of Contents for RT-5100
Page 12: ...X 161221687130 承認済 Uncontrolled copy when printed ...
Page 18: ... XVI 161221687130 承認済 Uncontrolled copy when printed ...
Page 46: ...28 BEFORE USE Getting Started and Exiting 161221687130 承認済 Uncontrolled copy when printed ...
Page 274: ...256 MAINTENANCE List of Replacement Parts 161221687130 承認済 Uncontrolled copy when printed ...
Page 302: ...284 EMC ELECTROMAGNETIC COMPATIBILITY 161221687130 承認済 Uncontrolled copy when printed ...