140
ADVANCED FEATURES
Parameter Settings
Program A: Fix, User
Factory setting: Fix
Selection of whether to use the default Program A or user-programmed one. Set the parameter to User
when Program A is reprogrammed. The default Program A can be used by resetting to Fix and press-
ing the Clear button or turning on the power again.
Program B: Fix, User
Factory setting: Fix
Selection of whether to use the default Program B or user-programmed one. Set the parameter to User
when Program B is reprogrammed. The default Program B can be used by resetting to Fix and press-
ing the Clear button or turning on the power again.
Program
: Fix, User
Factory setting: Fix
Selection of whether to use the default Program
or user-programmed one. Set the parameter to
User when Program
is reprogrammed. The default Program B can be used by resetting to Fix
and pressing the Clear button or turning on the power again.
Program #7 (For Plus Package): Fix, User
Factory setting: Fix
Selection of whether to use the default Program #7 or user-programmed one. Set the parameter to
User when Program #7 is reprogrammed. The default Program #7 can be used by resetting to Fix and
pressing the Clear button or turning on the power again.
Adjust ADD Subj
Final: Yes, No
Factory setting: Yes
Selection of whether or not to set the additional power taken into consideration of the difference of
between the subjective data and prescription S and C data in the prescription data automatically when
is pressed in the final field.
When there is no additional power data in the subjective field, this setting is disabled.
The data copying from Subj to Final
S/C/A, S/C/A/VA/ADD, S/C/A/VA/ADD/Prism, All data
Factory setting: S/C/A
Setting for the contents to be copied when copying the data from Subj to Final. The contents of copying
is only for the Far point mode.
Preset power of Subj 1: AR, LM
Factory setting: AR
When both AR and LM data are input, the refractor head is initially set for subjective refinement for the
selected setting.
The PD in the AR data is prioritized for use.
When only one of the two data is imported, the refractor is set with the imported data.
161221687130 承認済
Uncontrolled copy when printed.
Summary of Contents for RT-5100
Page 12: ...X 161221687130 承認済 Uncontrolled copy when printed ...
Page 18: ... XVI 161221687130 承認済 Uncontrolled copy when printed ...
Page 46: ...28 BEFORE USE Getting Started and Exiting 161221687130 承認済 Uncontrolled copy when printed ...
Page 274: ...256 MAINTENANCE List of Replacement Parts 161221687130 承認済 Uncontrolled copy when printed ...
Page 302: ...284 EMC ELECTROMAGNETIC COMPATIBILITY 161221687130 承認済 Uncontrolled copy when printed ...