Pacer Input
(Load dependent):
Amplitude (50 ohm): 0 - 10 mV
Resolution (50 ohm): 25
μV
Amplitude: (500 ohm): 0 - 100 mV
Resolution: (500 ohm): 1 mV
Accuracy:
±2%
Immunity Test:
Interference Signal: 50Hz / 60 Hz
ECG signal Amplitude:0.5 to 5.0mV
Refractory Period Measurement:
20 to 500 ms (both Pacing and Sensing)
Accuracy: ± 5 ms
Patient Simulator
Waveform Simulation: ECG (NSR),
ECG Rate: 30, 40, 50, 60, 70, 80, 90, 100, 110, 120, 130, 140,
150, 160, 170, 180, 190, 200, 220, 240, 260, 280, 300, 310
BPM
Accuracy: ± 1%
Amplitude: 0.1, 0.25, 0.5, 1.0, 1.5, 2.0, 2.5, 3.0, 3.5, 4.0, 4.5,
5.0 mV.
Accuracy: ± 2% @ Lead II (Square Wave)
Output Impedance: RL to all leads 1000 ohms.
High Level Output: 0.5 V/mV of low level (lead II)
Arrhythmia Selections:
Ventricular Fibrillation Coarse (VFBC)
Ventricular Fibrillation Fine (VFBF)
Ventricular Tachycardia (VTAC)
Atrial Fibrillation (AFIB)