Waveform storage and playback function.
12 lead ECG output.
7.4 Volt LI Ion batteries.
1M internal flash memory for storing test records.
Rs 232 interface for uploading test data.
Intended Use
The Delta 3300 is indented to be used to determine the performance
specifications of the defibrillators and transcutaneous pacemakers
Specifications
Defibrillator Analyzer
Energy Output Measurement:
Maximum Energy: 800 J
Display Resolution
:
Based on
Auto Range
Auto Range: 2 - 9.99 J, 10-99.9 J, 100-800 J
Resolution:: 0.01 J ( 2 -9.99), 0.1 J (10-99.9), 1 J (100-500)
Accuracy: ±1 % of reading ± 0.1 J (2-9.9), ± 1J (10-99), ± 2 %
(100-500)
Load Resistance:
50 ohms ±1 %; non-
inductive (< 1 μH)
Sampling Time: 20 uS
Measurement Time Window: 60 ms
Absolute Max Peak Voltage: 5000 V
Absolute Max Peak Current: 100 A
Pulse Width
:
60 ms
Waveform Storage / Playback:
Discharge waveform: ECG outputs leads and through paddle
contacts.
High Level Output
:
100:1 Time Base Expansion