Charge Time Test
The charge time function is used to test the charging time of the
capacitor and the battery of the defibrillator.
1. Set the defibrillator energy to maximum (
do not charge
at this time
).
2. Select “CTime (F2) from the menu.
3. Select “Start” (F4).
4. Set the desired count down time. The countdown starts
from the set value and shows on the screen.
5. Place the paddle firmly on the contact plates and wait
for the beep.
6. At the beep press the charge button on the paddle of the
defibrillator and press the discharge button of the
defibrillator.
7. When the defibrillator is fully charged it automatically
discharges into the paddle and the time will be displayed on
the screen.