Sync Time Measurements:
Timing Window: Starts at each R wave peak
Test Waveforms: User selectable ECG rates.
Delay Time Accuracy: ±1 ms.
Charge Time Measurement:
Countdown Timer: User selectable - 30, 45, and 60 seconds.
Energy Charge Time: Display the charge time measured from
the time the start button is pressed to the discharge.
Transcutaneous Pacemaker
Test Load Range
:
50 to 1100 ohms in step of: 100 ohms.
Accuracy:
50 - 500 ohm ±1 %, 500 - 1100 ohm ±1.5 %
Pulse Measurements:
Amplitude: 4 to 400 mA (50 ohm load)
Accuracy: ± 1 % or ± 1mA
Max. Amplitude: 400 mA all loads
Rate: 30 to 800 ppm
Accuracy: ±1 % or 2 ppm
Pulse Width: 0.5 to 50 ms
Accuracy: ±1 % or ±0.3 ms
Demand Sensitivity Test:
Waveforms: QRS, Square (SQR), Triangle (TRI), and
Haversine (SSQ)
Waveform width:
40, 50, 60, 70, 80, 90, 100, 200, 300, 400 ms
Pacer Rate:
30 to 120 ppm
ECG Output:
Amplitude: User selectable: 0.1, 0.25, 0.5, 1.0, 1.5, 2.0, 2.5, 3.0,
3.5, 4.0 and 5 mV