9 Testing with Zodiac Diagnostic or Clinical - Stand-Alone
47
MADSEN Zodiac
The detected reflex is shown in the results area.
•
Stimulus type level
•
Progress indicator during measurements
•
Deflection curves are shown when a reflex has been detected.
If a reflex is not detected, the highest stimulus is displayed together with a
No Response
indication.
If the system cannot automatically analyze a reflex measurement, a
Not Tested
indication is shown.
C. Reflex Screening results
The results table shows the results related to the currently selected curve:
•
Probe
(the probe tone in Hz)
•
TPP
(Tympanometric Peak Pressure)
•
ECV
(Equivalent Ear Canal Volume)
•
SA
(Static Peak Admittance), or
SC
(Static Peak Compliance) when volume equivalent units are used
•
TW/Ratio
(Tympanometric Width/Tympanometric Ratio). Describes the steepness of the curve.
•
Type
(Jerger types A, As, Ad, B, C, D and E denote the shape of the 226 Hz curve). You can set the type to be
determined automatically, and you can subsequently change it manually, if necessary.
D. Tympanometry results
E. On-screen selections
Curve selection
Press the
Previous Curve
or
Next Curve
soft button to select the curve in the
Reflex Screening and Tympanometry result tables.
Manually adjusting the TPP
Press the
Edit TPP
soft button, use the dial to move the marker to the TPP,
and press the
Done
soft button.
Changing tympanogram curve type Press the
Edit TPP
soft button. Press the
Change Type
soft button until the
type is shown and press the
Done
soft button.
Deleting a curve
Select the curve you wish to delete and press the
Delete
button.
To replace a single measurement, delete a curve and adjust the measurement
selections to redo the single measurement.
The graph area shows the tympanometric curves and can rescale automatically to fit the curves.
•
Tympanometric curves
•
Pressure and admittance scales
•
Ear canal volume bar. The ear canal volume is shown to the right of the graph.
•
Norm area
B. Tympanometry graph