8 Testing with Zodiac Quick Check - Stand-Alone
42
MADSEN Zodiac
E. On-screen selections
Curve selection
Press the
Previous Curve
or
Next Curve
soft button to select the desired curve in
the Reflex Screening and Tympanometry result tables.
Manually adjusting the TPP
Press the
Edit TPP
soft button, use the dial to move the marker to the desired
TPP, and press the
Done
soft button.
Changing tympanogram curve type Press the
Edit TPP
soft button. Press the
Change Type
soft button until the
desired type is shown and press the
Done
soft button.
Deleting a curve
Select the curve to delete and press the
Delete
button.
To replace a single measurement, delete a curve and adjust the measurement
selections to redo the single measurement.
The results table shows the results related to the currently selected curve:
•
Probe
(the probe tone in Hz)
•
TPP
(Tympanometric Peak Pressure)
•
ECV
(Equivalent Ear Canal Volume)
•
SA
(Static Peak Admittance), or
SC
(Static Peak Compliance) when volume equivalent units are used
•
TW/Ratio
(Tympanometric Width/Tympanometric Ratio). Describes the steepness of the curve.
•
Type
(Jerger types A, As, Ad, B, C, D and E denote the shape of the 226 Hz curve). You can set the type to be
determined automatically, and you can later change it manually, if necessary.
D. Tympanometry results