6.5
Madsen A450 connection panel
Install Otosuite on the PC before you connect Madsen A450 to the PC.
The installation must be carried out in accordance with IEC 60601-1 (Ed. 3.1), UL60601-1, and CAN/CSA C22.2
No. 60601-1-14:2014.
It is a general rule for all electrical equipment used in the proximity of the client that:
•
The connected equipment must comply with IEC 60601-1 (Ed. 3.1) except for the PC.
See also
When you have connected the accessories, configure your system setup. To do so, see
Configuring the Audiometry Module
1. PC/USB connection
2. External power supply
3. Sound field speakers (power output)
4. Insert Earphones
5. Headphones - air conduction
6. Patient Responder
7. Speaker, Analog (line output)
8. Operator monitor headset - headphones
9. Operator monitor headset - boom microphone
10. Insert Earphones
11. Headphones - air conduction
12. Bone conductor
13. Line-in
14. Counseling and Simulations headphones
15. Talk-back microphone
Note
•
Blue corresponds to Left and red corresponds to Right.