20
Function
Description
2
Fuse
F1 and F2 Mains
9031D040x: 110V/120V: T10A / 250V~
9031D041x: 200/240V: T6.3A / 250V~
3
M4 Thread for Electrical Enclosure Contact and Additional Protective Earth.
4
Potential equalization
Where potential equalization is necessary or desirable, a potential equalization conductor may be connected to this
connector as per DIN 42 801.
5
Spring Retaining Clamp for Power Cord
Place the spring retaining clamp around the power cord to ensure a safe connection.
6
Auxiliary Outlet
-
6 Outlets: 50/60Hz
9031D040x:
100-120V~ total max.
- Continuous Operation 7.4A
- Momentary Operation 12.4A
7.4A equals 850VA, and 12.4A equals 1425VA at 115V~
9031D041x:
200-240V~ total max.
- Continuous Operation: 3.7A
- Momentary Operation: 6.2A
3.7A equals 850VA and 6.2A equals 1425VA at 230V~
7
SN: Device Serial Number
8
REF: Device Reference Number
NOTE
: Unplugging the power line cable from the mains input on the Isolating Transformer disconnects the mains power to all supplied
devices.
Isolation Transformer Classification
Type of protection against electric shock:
•
Class I: Equipment in which protection against electric shock does not rely on basic insulation only, but which includes an additional
safety precaution in that means are provided for the connection of the equipment to the protective earth conductor in the fixed
wiring of the installation in such a way that accessible metal parts cannot become live in the event of a failure of the basic insulation.
Method(s) of sterilization:
•
Device is not intended to be sterilized.
Degree of patient vicinity:
•
Device suitable for use in patient environment.
Degree of protection against electric shock:
•
Enclosure protectively earthed (treated as applied part Type B).
Degree of protection against harmful ingress of water:
•
IP20: Ordinary equipment (enclosed equipment without protection against ingress of water).
Degree of safety of application in the presence of a flammable anesthetic mixture with air or with oxygen or nitrous oxide:
•
Equipment not suitable for use in the presence of such a mixture.
Mode of operation:
•
Continuous operation at normal load and Momentary operation at high load.
•
Momentary operation at high load is used for short time printing with high power laser printers.
•
Momentary operation at high load is possible for up to 60 sec per hour at elevated temperatures.
Operation at high load over prolonged time may cause thermal cut out. After cooling down (hours) internal thermal fuse will reset and normal
operation can be resumed.