41
Nerve Conduction Data Page
Activating the Function Key
NC Data
button moves to a summary page, displaying the result table for all Nerve conduction data from the
current study. Clicking this button again or closing the summary page window advance back to the Test page.
Settings and Design
The Settings Page
The above display
is a Settings example for NCS
From the
Current Settings
page, the recording and stimulation parameters of the test can be adjusted.
•
Click on the anatomy set to view the settings of a site.
•
To open the site’s settings of a specific trace from the scope window, right- click on the trace, and select Settings.
•
After adjusting the settings, click OK to save the settings or click
Cancel
to quit the settings page without saving the changes.
The customized settings carried out through Settings and, if available, Design pages are only valid for that specific test and patient. New tests
will use the default test settings.
Reassigning Traces in Motor NCS
Sweeps acquired on the wrong stimulation site or wrong nerve may be reassigned.
•
Right-click on a stimulation site label to access reassign function.
•
Top half of the options reassign the selected trace to other stimulation sites for the same selected nerve.
•
Bottom half of the options reassign all traces from this selected nerve to another nerve present in the test menu.
Sweeps acquired on the wrong side of the patient may be reassigned.
•
Right-click in the sweep window background to access the
swap side
function
•
All nerve data recorded on the left side of the patient are reassigned to the right side, while all data recorded on the right side are
reassigned to the left side.