Maintenance
32
| BRM2 Brain Monitor Reference Manual
Cleaning
The components of the BRM2 Brain Monitor are designed to withstand normal hospital cleaning
protocols, however they must not be subjected to any form of sterilization procedure. Please note the
following cleaning instructions and warnings.
Monitor (including touchscreen) and Isolation Unit
Clean the Monitor and Isolation Unit by wiping with a soft cloth dampened with detergent and water.
Data Acquisition Unit and connecting cables
Normal cleaning:
Clean by wiping with a soft cloth dampened with detergent and water.
After use on an infected baby:
Clean by wiping with a soft cloth dampened with 70% v/v isopropyl
alcohol (IPA). Allow to dry naturally.
Disconnect the BRM2 Brain Monitor from the AC supply outlet before cleaning. Use only a
damp cloth for cleaning (do not use liquid or spray detergents).
Never pour liquid into any opening in any part of the BRM2 Brain Monitor, or submerge
any component in liquid. This may cause fire or electrical shock.
Do not autoclave any component of the BRM2 Brain Monitor, as serious and irreversible damage may
result. Do not clean with concentrated bleach, corrosive chemicals or abrasive cleaning compounds.
The Monitor, Data Acquisition Unit and Isolation Unit must not be opened except when being serviced
by an Authorized Service Center, or when following written instructions from BrainZ.
The surface of the Monitor touchscreen is resistant to the following substances (at room temperature):
Acetone
Methylene chloride
Isopropyl alcohol
Hexane
Ammonia-based glass cleaners
Turpentine
Mineral spirits.