Brainz BRM3 Brain Monitor Reference Manual |
31
Maintenance
Checking for normal operation
The BRM3 Brain Monitor is supplied with a test plug, attached to the DAU. The test plug can be used to
terminate each EEG channel for test purposes. This procedure constitutes a functional test of the left and
right EEG channels.
This check should be performed when the BRM3 Brain Monitor is first assembled and as part of the
annual maintenance check. See
Checking for normal operation
on page 19.
Message codes
For a list of message codes and their explanations, see the
BRM Onscreen Help System
or the
BRM3 Brain Monitor Technical Manual.
Cleaning
The components of the BRM3 Brain Monitor are designed to withstand normal hospital cleaning
protocols, however they must not be subjected to any form of sterilization procedure. Note the following
cleaning instructions and warnings.
BRM3 Brain Monitor (including Touchscreen Monitor), DAU and connecting cables
Normal cleaning:
Clean by wiping with a soft cloth dampened with detergent and water.
After use on an infected baby:
Clean by wiping with a soft cloth dampened with 70% v/v isopropyl
alcohol (IPA). Allow to dry naturally.
Disconnect the BRM3 Brain Monitor and the printer (if fitted) from the AC supply outlet
before cleaning. Use only a damp cloth for cleaning. Do not use liquid or spray
detergents.
Never pour liquid into any opening in any part of the BRM3 Brain Monitor, or submerge
any component in liquid. This may cause fire or electrical shock.
Do not autoclave any component of the BRM3 Brain Monitor, as serious and irreversible damage may
result. Do not clean with concentrated bleach, corrosive chemicals or abrasive cleaning compounds.
The Touchscreen Monitor and DAU must not be opened except when being serviced by an Authorized
Service Center, or when following written instructions from Natus.