CF46 IP Address Subnet Mask Prefix Setting
If you want to set the IP address manually (MANUAL IP function), set the subnet mask prefix.
This parameter is displayed and can be set only when the LAN interface is in use.
This parameter is applied when you turn the POWER switch off and on or when you execute
APPL with CF40.
CF47 SCPI-RAW Protocol TCP Port Number Setting
Set the TCP port number of the SCPI-RAW protocol for when the LAN interface is in use.
This parameter is applied when you turn the POWER switch off and on or when you execute
CF40. TCP port number 4880 cannot be used.
Parameter Number
Value
Description
CF46
16 to 31
Sets the subnet mask prefix.
Examples:
Display example when “16” is set
255.255.0.0
Display example when “24” is set
255.255.255.0
Display example when “31” is set
255.255.255.254
Value
Parameter Description
4000 to 9999
(except 4880, 5024, and 5044)