Connecting mmRH-3647/3657 Radio Heads to a Bidirectional System
Complete the following steps to connect mmRH-3647/3657 mmWave radio heads to a
bidirectional mmWave Transceiver System.
Note
Connect the modules of the mmWave Transceiver System for standard use
before connecting mmRH-3647/3657 radio heads.
1.
Connect to the DIGITAL I/O connector of the PXIe-3620 front panel using the joint end
of a dual-channel EPLSP cable.
2.
Connect one end of the dual-channel EPLSP cable to the CONTROL connector of the
mmRH-3657 back panel.
3.
Connect the other end of the dual-channel EPLSP cable to the CONTROL connector of
the mmRH-3647 back panel.
4.
Connect the RX LO1 connector of the PXIe-3620 front panel to the LO IN connector of
the mmRH-3657 back panel using an SMA(m)-to-SMA(m) cable.
5.
Connect the RX IF IN connector of the PXIe-3620 front panel to the IF OUT connector
of the mmRH-3657 back panel using an SMA(m)-to-SMA(m) cable.
6.
Connect the TX LO1 connector of the PXIe-3620 front panel to the LO IN connector of
the mmRH-3647 back panel using an SMA(m)-to-SMA(m) cable.
7.
Connect the TX IF OUT connector of the PXIe-3620 front panel to the IF IN connector of
the mmRH-3647 back panel using an SMA(m)-to-SMA(m) cable.