Table 14. Absolute Amplitude Accuracy (Continued)
Frequency
Temperature 23 °C ±5 °C (dB)
3 GHz to 4.4 GHz
1.8
Note
Absolute amplitude accuracy uses a correction coefficient in EEPROM to
improve performance. The TX amplitude accuracy applies to the output power
level from -12 dBm to -4 dBm.
Figure 11. TX Output Power
–1.0
–500.0 m
500.0 m
0.0
1.0
1.5
2.5
2.0
3.0
3.5
4.5
4.0
5.0
5.5
6.0
6.5
7.5
7.0
8.0
9.0
8.5
9.5
10.0
1.0 G
1.5 G
2.0 G
2.5 G
3.0 G
3.5 G
4.0 G 4.4 G
200.0 M
Frequency (Hz)
TX Output P
ow
er (dBm)
Noise Density
Note
All values are typical.
Table 15. Noise Density
Frequency
Temperature 23 °C ±5 °C (dBm/Hz)
>200 MHz to 1 GHz
-138
>1 GHz to 2 GHz
-138