NGSLR Operations Manual
Performing MINICO and Stability Tests
Document Number:
P a g e |
4 8
NASA NGSLR OPS Manual (v1.0)
Figure 6 7: Schedule control window
16.
Collect data from each calibration pier in this order: (C A B C A B
)
a. On RAT, go to the
Control
=>
Schedule
submenu (Figure 6 8).
b. Open the
Cal Targets
pull down menu and select the next target (Figure 6 8).
d. Adjust the ND value under the
Misc Control
menu as determined previously (Figure 6 3).
c. Ensure data is being collected Go to the
Misc Control
menu and from the
Data Logging
pull
down menu, set
All Returns
and select
Apply
.
d. Continue collecting returns for ten minutes, or the specified amount of time.
e. Record the ND value used for the current target on the log sheet.
f. Repeat step 16 until data has been collected for calibration piers in this order (C A B C A B C).
Figure 6 8: Schedule control window
17.
Exit POP to stop collecting returns
To do this, go to the DAM/POP console and type
x <ENTER>