84
|
2017-07 MAN-C0034-06
USER MANUAL
nCounter
®
Dx Analysis System
19.
To release the fixture after the run is complete, pull the latch on the top of the device up and towards the front of the system using a
finger as shown in
FIGURE 5.29
.
FIGURE 5.29:
Releasing the fixture after a completed run
20.
After processing is complete, it is important to do the following:
a.
Seal the wells immediately with the adhesive cover provided to prevent evaporation.
b.
Protect Samples from as much light as possible.
c.
If the Cartridge will not be scanned on the Digital Analyzer within an hour the sealed Cartridge should be stored at 4°C in an
opaque box. The Cartridge can be stored this way for up to one week with minimal degradation.
d.
Empty the waste containers.
21.
Touch
Finish
to return to the Main Menu.