78
|
2017-07 MAN-C0034-06
USER MANUAL
nCounter
®
Dx Analysis System
14.
Carefully lower the electrode fixture in place over the Cartridge (
FIGURE 5.20
). The 24 electrodes should insert easily into the 24
wells. Touch
Next
.
FIGURE 5.20:
Carefully lower the electrode fixture into place over the Cartridge.
IMPORTANT:
Do not use the release handle to lower the fixture. Doing so will prevent the fixture from locking. Instead, press
on the body of the fixture away from the release handle (
FIGURE 5.20
).
IMPORTANT:
IIf any resistance is felt while lowering the fixture, stop and adjust the position of the Cartridge slightly. Make sure
that the electrodes are correctly aligned. If the electrodes are not aligned, re-align the electrodes using the ‘Align Electrodes’
workflow under the Maintenance Menu. The Prep Station will not be able to process any of the samples if electrodes are bent.