76
|
2017-07 MAN-C0034-06
USER MANUAL
nCounter
®
Dx Analysis System
11.
Place the tip sheaths on the deck and press firmly into place. Touch
Next
.
>>> The ‘Empty Strip Tubes’ screen will appear (
FIGURE 5.16
).
FIGURE 5.16:
The ‘Empty Strip Tubes’ screen
12.
Place the empty strip tubes in the heater on the deck. Touch
Next
.
>>> The ‘Scan Sample Cartridge’ screen will appear.
NOTE:
Only one empty heater strip tube is required for runs performed using a 1-, 2-, 3-, or 4-test kit. For these Run Sets, load
the empty heater strip tube into the front position of the heat block (nearest the user) on the Prep Station deck.
FIGURE 5.17:
The ‘Scan Sample Cartridge’ screen