User's Manual

 of GNF001 

 

Version 1, July 2023 

- 5 - 

non-compliance with the following safety precautions happens, the centrifuges would 

cause the operator or other personnel injury or make damages to centrifuge separation 

plant and the internal sample: 

Centrifuge design is neither anti-corrosion, nor is proof, therefore, ensure no 

centrifuges corrosion in the environment and the possibility of an explosion happen in 

the use of the environment; 

The use of centrifuges is strictly prohibited in the following materials:   

- Flammable and explosive materials;   

- Strengthen the role of materials science;   

- Toxic or radioactive substances or pathogenic micro-organisms and so on. 

For the separation of corrosive substances and pathogenic micro-organisms cell 

effective sealing measures should be carried out in advance and effective disinfection 

measures should be carried out after use. Details, see “repair and maintenance issues - 

sterilization” in the contents. 

The substances for the separation of corrosive can cause damage and destruction of 

centrifuge internal material or weaken the rotor mechanical strength, so when 

separate of corrosive substances , the corrosive substances must be kept in the 

protective container.   

 

1.1

 Operation Notes

 

Before centrifuge run (sample separation), it must be confirmed with appropriate 

rotor and ensure a solid installation. 

When in the process of running centrifuges ( rotors turning) or in the process to 

stop centrifuge ( but when the rotor is still rotating), do not manually open the door or 

move the centrifuge;   

The components used in The company. low-speed Benchtop Air-refrigerate 

Centrifuge must be dedicated accessories provided byThe company... A number of 

common components, such as: glass and plastic containers used for the separation, it 

should be confirmed to meet the requirements to use the products, which shall 

conform to the corresponding rotor speed and the largest of the greatest use of 

centrifugal force, etc. 

Do not open the door of the cases when using the centrifuges or in separation of 

sample; 

The replacement of the mechanical parts of the centrifuge and electronic devices, 

must be implemented by relevant personnel designated by theThe company..; 

When the operator uses the centrifuges, it is important to choose an appropriate 

load of the rotor, and shall not overload the use of the rotor; 

Ensure regular checks of the rotor, if the rotor has obvious corrosion or obvious 

signs of damage, using must be stoped; 

After using for a period of time, the provisions of maintenance should be strictly in 

accordance with the "cleansing and disinfection," 

1.2

 The relevant quoting standards

 

Summary of Contents for GNF001

Page 1: ...eed refrigerated centrifuge safety rules For safety reasons please read the operating manual carefully before starting the operation Components exposed to heat humidity and biological changes can caus...

Page 2: ...machines the centrifuge may be damaged by following factors Chemical effects Environmental impacts including radiation of natural UV The corrosion and abrasion and wear of shield and other security c...

Page 3: ...efrigerate Centrifuge must be conducted at the specified time intervals Security Warning Signs Description Caution Before using of equipment please read the Manual carefully Caution low pressure dange...

Page 4: ...tion 10 4 3 Fixed placement machine 10 4 4 Connect the power supply correctly 10 5 Operating Instructions 11 5 1 Introduction to Control Panel and Display Interface 11 5 2 Power on 12 5 3 open 12 5 4...

Page 5: ...esign Not pay attention to or understand the rules of safe use Any person who involving the using or maintenance of The company low speed BenchtopAir refrigerate Centrifuge should read and understand...

Page 6: ...ner 1 1 Operation Notes Before centrifuge run sample separation it must be confirmed with appropriate rotor and ensure a solid installation When in the process of running centrifuges rotors turning or...

Page 7: ...afety requirements GB4793 7 2001 Safety requirement of measurement and control of laboratories electrical equipment special requirements of laboratories electrical centrifuge GB191 2000 Icon of packag...

Page 8: ...low speed freezing centrifuge 2 1 GNF001 Introduction to the Appearance of Low Speed Refrigeration Centrifuge Figure 1 Left Side Appearance of GNF001 Low Speed Refrigeration Centrifuge Figure 2 Exter...

Page 9: ...components 2 3 1 The refrigeration system includes compressors condensers fans etc Adopting a powerful fluorine free high efficiency refrigerant R134a the temperature control range is between 20 and...

Page 10: ...ludes settings for rotational speed and centrifugal force operating time selection of lifting and lowering rate control of the entire machine display system and alarm system To ensure the normal opera...

Page 11: ...he machine will stop running in a timely manner and issue an alarm prompt It can only be restarted after cooling down in the centrifugal chamber Emergency opening During the operation of the rotor if...

Page 12: ...our company regarding the technical specifications and purchase number of the optional auxiliary parts Table 1 Rotor type and technical parameters Rotor capacity ml tubes Max spee d r min Max Rcf g Te...

Page 13: ...support or table for placing the centrifuge should be firm and not shake or vibrate If a movable bracket or trolley is used one with a locking device should be used To ensure the safe operation of the...

Page 14: ...orrectly to the machine power socket and connect it firmly to the network power supply When the power switch is closed it is and when the power switch is disconnected it is O 5 Operating Instructions...

Page 15: ...on the power switch located on the right side of the machine The LCD color display on the control panel lights up After the self inspection of the machine is completed enter the Home screen and now yo...

Page 16: ...e rotor types and centrifuge tube specifications and other optional components can lead to poor centrifuge performance and even damage to the centrifuge The steps for installing the rotor are as follo...

Page 17: ...r body Figure 6 schematic diagram of installing the cradle After installing the rotor before each use or after a period of use the installation position of the rotor should be checked for any changes...

Page 18: ...le load maximum allowable load Actual load actual load Do not overload the rotor otherwise it will cause the rotor to explode and the debris generated by the explosion will damage the centrifuge 5 7 F...

Page 19: ...ust be tightened before starting and running 5 8 5 Centrifugal test tubes can load samples at different times but they must be symmetrically loaded with a permissible weight error of 1 5g When loading...

Page 20: ...meter adjustment knob key can be pressed vertically downwards or the system will automatically flash three times as the default 5 9 6 ACCEL DECEL speed up and down setting the acceleration and deceler...

Page 21: ...ing if you need to stop halfway please press the key The set time will gradually decrease from the set value to zero When the time value shows zero the centrifuge will automatically stop and the speed...

Page 22: ...ain After resetting the parameters there is no need to manually confirm and the system will automatically flash three times to confirm the settings Before setting parameters the rotor must be correctl...

Page 23: ...ord The regular or based on usage cleaning and maintenance tasks mainly involve the centrifuge casing inner chamber rotor and separation containers used This is to prevent pollutants from being left o...

Page 24: ...ay be carried away and cause corrosion Allow to dry upside down after cleaning 6 1 3 5 Use a cloth or tweezers to remove debris from the centrifuge chamber 6 1 3 6 The connecting parts of the motor sh...

Page 25: ...rsonnel are responsible for free on site repair If the replacement parts are required due to their own quality rather than improper use they can be replaced for free otherwise they must be paid The sh...

Page 26: ...ing Please be patient The steps for emergency door opening are as follows 7 1 1 Confirm that the rotor has completely stopped 7 1 2 Turn off the power switch 7 1 3 Using the accompanying tool insert i...

Page 27: ...replaced the accessories Error 4 Hall sensor failure Please contact us to change the motor Error 5 The brake overvoltage Please check the brake resistance was connected well or the brake resistance wa...

Page 28: ...l Heavy vibration of machine 1 The following situation is normal The speed of rotor is over the critical speed and the machine vibrates 2 Check whether the rotor is locked tightly or not 3 Check the s...

Page 29: ...using time of the machine and accumulate data of operation of the machine to do the diagnostic or get the reason of accident please record the running of the operation every time Recommended Record L...

Page 30: ...60HZ 10A Set time range 1 99 hours 1 99 minutes 1 59 seconds Three modes are available for selection Accuracy 1 second Maximum speed 6000rpm Maximum relative centrifugal force 5150 g Maximum capacity...

Page 31: ...line 1 Provided by country region 3 Root Supply according to the order contract 4 Special Hex key 1 5 20mllube 1 6 certificate of compliance 1 7 Operation Maintenance and Repair Manual 1 When unpackin...