6.1.3 maintenance
6.1.3.1
Do not use sharp objects to collide with the rotor. During transportation and
disassembly, it is necessary to prevent collisions and prevent cracks in the rotor during
use due to scratches or external injuries.
6.1.3.2
Regularly inspect the rotor components (especially the bottom of the test tube
hole) for corrosion spots, grooves, and small cracks. If any of the above conditions are
found, please stop using the rotor and contact our company.
When disassembling the rotor, grab it with both hands and lift it
vertically, do not shake it left or right!
6.1.3.3
Normally, the rotor is cleaned once a week. If separating salt solutions or other
corrosive samples, please clean them immediately after use. If the sample is found
splashing, soaking, or dripping on the rotor during use, it should be immediately
sucked dry and locally cleaned.
6.1.3.4
When cleaning the rotor, please use neutral detergent to wet sponge or cotton
cloth, and then use distilled water to wash the detergent. Do not sprinkle or spray the
rotor with water, as the liquid may be carried away and cause corrosion. Allow to dry
upside down after cleaning.
6.1.3.5
Use a cloth or tweezers to remove debris from the centrifuge chamber.
6.1.3.6
The connecting parts of the motor shaft and rotor shaft hole should be coated
with lubricating grease.
6.1.3.7
Steps for maintaining the motor spindle:
●
Turn on the power switch and wait for the self check to complete;
●
Press the
key to open the centrifuge door cover;
●
Using the special tool provided with the machine for disassembling and
disassembling the rotor, loosen the locking sleeve and take out the rotor. Clean the
locking sleeve. Note that the clockwise direction is for tightening the locking
sleeve, and the counterclockwise direction is for loosening the locking sleeve;
●
Clean the conical surface of the motor spindle without leaving any dirt. Add an
appropriate amount of lubricating oil or apply it with lubricating paper;
6.1.3.8
When disassembling the machine, the power must be cut off first and the
power cord connected to the back wall of the machine must be unplugged. It is not
allowed to operate with electricity to prevent personnel from getting electric shock or
annex
Max Temp
(℃)
Minimum time
(
min
)
Maximum time
(
min
)
Maximum
number of times
glass
tube
134
—
138
3
5
—
PCtube
115
—
118
30
40
20
PPtube
115
—
118
30
40
30
PAtube
115
—
118
30
40
20