After setting, it is shown in the following figure:
Figure 7: Example parameter setting completed
5.10 Parameter settings Other
5.10.1 RcfSettings
:
Press the
key twice continuously on the control
panel, causing the number at Rcfto flash
→
turn the parameter adjustment
knob key to set its value. Note: The Speed value is automatically converted
with the Rcm value;
5.10.2
During the parameter setting process, if an alarm occurs due to a machine
malfunction or incorrect parameter settings, press the
key to cancel the
alarm and then refer to the rotor number again for setting.
5.10.3
Press the
key and the machine will start running (if you need
to stop halfway, please press the
key). The set time will
gradually decrease from the set value to zero. When the time value shows
zero, the centrifuge will automatically stop, and the speed will gradually
decrease from the set value to zero (the time required for the speed to
decrease from the set value to zero is related to the speed reduction
setting). When the speed becomes zero, the door cover will automatically
open and the machine will emit shutdown music, Press the
key to
stop the sound. Centrifuge completion.
5.10.4
If instantaneous centrifugation is required:
Please press and hold
the
key on the control panel, and the speed will continue to
increase. Release the key to stop, and the maximum speed during this time
will be the set speed corresponding to the rotor number and increase in
Sec. unit time.
5.10.5
提示音设置:
When the instrument is in standby mode, both the