To save EDID data to the tester internal memory, tap the
Save
button.
To download EDID data, connect the USB memory stick and tap the
Download
button.
Two files will be saved
–
EDIDXXXX.txt
and
EDIDXXXX.bin
.
4.
To delete EDID data from tester, select the files to delete and tap the
Delete
button.
4.2.4.3
Certification Report
The certification report menu provides the following capabilities:
◼
Run Certification
–
report data generation
◼
Report Files
–
saves the report data from the tester to the USB memory stick
◼
Cable Profile
–
Cable profile configuration
4.2.4.3.1
Run Certification
The HDBaseT report data collection feature is used to collect the data required to perform HDBaseT
cable analysis.
To access data for HDBaseT cable analysis:
Connect the
MU-M4SOL
test device to both ends of the CATx cable, as shown below in
◼
The HDMI cable
must
be connected to the Internal Generator to ensure that HDMI
traffic is running on the HDBaseT link.
◼
The “local” side of the cable is
connected to either the transmitter or receiver module,
housed in the
MU-M4SOL
chassis.
◼
The “remote” side of the cable is
connected to the other
MU-M4SOL
module.
Make sure
that the remote module is terminated by a Pluggable POH Adapter as
shown below.
Figure 33: Cable Test Setup
HDBaseT
HDMI
Summary of Contents for MU-M4SOL
Page 1: ...Test Device Model MU M4SOL User s Manual Revision A16 ...
Page 39: ...MU M4SOL User s Manual Page 39 of 130 Figure 45 Log Report ...
Page 68: ...MU M4SOL User s Manual Page 68 of 130 Figure 83 Information Menu ...
Page 108: ...MU M4SOL User s Manual Page 108 of 130 Figure 134 IP Module report ...