![multicomp pro MP720105 Manual Download Page 51](http://html1.mh-extra.com/html/multicomp-pro/mp720105/mp720105_manual_1854300051.webp)
4.Advanced User Guidebook
of the screen, for example,
,
indicates that trigger type is
windows, trigger source is CH1, polarity is positive, 0.00mV the differential between
up level and low level threshold.
Windows Trigger menu
list:
MENU
SETTING
INSTRUCTION
Single
Mode
Windows
Set vertical channel trigger type as Windows trigger.
Source
CH1
CH2
Select CH1 as the trigger source.
Select CH2 as the trigger source.
Threshold
Up Level
Low Level
Adjust the
M
knob or click
to set the up level
threshold.
Adjust the
M
knob or click
to set the low level
threshold.
Condition
Polarity
Positive Polarity, the oscilloscope triggers on the
positive Windows pulse.
Negative Polarity, the oscilloscope triggers on the
negative Windows pulse.
Enter: Triggers when the trigger signal enters the
specified trigger level range.
Exit: Triggers when the trigger signal exits the
specified trigger level range.
Time: Specify the hold time of the input signal after
entering the specified trigger level. The oscilloscope
triggers when the accumulated hold time is greater than
the windows time. Available range is 30ns-10s, default
100ns.
46
Summary of Contents for MP720105
Page 1: ...i Digital Storage Oscilloscope Model MP720105 MP72016 MP72107...