ModCon®75 Touch
•
Touch the white box the left of ‘Store Previous Tag Information’ to add check mark.
Checking this box will store all previous tag information when attempting to make
changes in case the user wishes to revert to a previous state.
•
X3 ID’s: to view network registry list.
•
Sensor Tag’s: to view sensors connected to an individual Ultima X3.
•
Edit ON: to edit X3 tag (X3 ID’s button) and edit sensor tag’s (Sensor Tags button).
•
Press Home - to go back to Main Menu Screen.
•
If new X3 transmitters/sensors are added/removed from the Modbus network, user
must run Start Scan to repopulate the network registry. If Start Scan is not run, the
output from added/removed X3 transmitters/sensors, including alarm/fault conditions,
will not be displayed on the MdoCon75 Touch Controller.
FAILURE TO FOLLOW THIS WARNING CAN RESULT IN SERIOUS PERSONAL
INJURY OR LOSS OF LIFE
Figure 30. Add Units to the Network
To add X3 transmitters/sensors to the network: press Start Scan once the following three
steps are completed:
1. First, verify all X3 transmitters/sensors (up to 25 X3 transmitters each with up to 3
sensors for a maximum of 75 sensors) are wired properly to the network.
2. Next, confirm each X3 transmitter connected to the network has a unique ID set by
using the Ultima/Ultima X Series Controller and Calibrator (Valid network ID’s are 2-
26).
NOTE
:
o
ID 0 is the broadcast Modbus address; ID 1 is the ModCon75 Touch address.
o
Baud rate = 19200 and Parity = even.
3. Press and hold the Start Scan button until the scanning begins, and button is greyed
out.
NOTE:
o
Verify that the ModCon75 Touch Controller found all Ultima X3
transmitters on the network. If the total number is less than the actual
number on the network, check your wiring and make sure that each
X3 transmitter on the network has its own unique address in the
range of 2 through 26.
This process may take several minutes.