When the pump inlet, sample line or probe is blocked, the pump alarm
must activate.
If the alarm does not activate:
1. Check the sample line and probe for leaks.
2. Once leak is fixed, recheck pump alarm by blocking the flow.
3. Press the ▲ button to reset the alarm and restart the pump.
Do not use the pump, sample line, or probe unless the pump
alarm activates when the flow is blocked. Lack of an alarm is
an indication that a sample may not be drawn to the sensors,
which could cause inaccurate readings.
Failure to follow the above can result in serious personal
injury or death.
Never let the end of the sampling line touch or go under any
liquid surface. If liquid is drawn into the instrument,
readings will be inaccurate and instrument could be
damaged. We recommend the use of an MSA sample probe
containing a special membrane filter, permeable to gas but
impermeable to water, to prevent such an occurrence.
During operation, a pump alarm may occur when the:
• Flow system is blocked
• Pump is inoperative
• Sample lines are attached or removed.
To Clear Pump Alarm
1. Correct any flow blockage.
2. Press the ▲ button.
• The Pump will now restart.
"
WARNING
4-3
Summary of Contents for altair 5
Page 68: ...6 3 Instrument Replacement Parts 6 3 ...
Page 71: ...7 Appendix Flow Charts 7 1 Start Up Sequence Power ON 7 1 ...
Page 72: ...7 2 Normal Operation 7 2 ...
Page 73: ...7 3 Options from Main Page 7 3 ...
Page 74: ...7 4 Password Protected Options 7 4 ...
Page 75: ...7 5 Menu Sequence from Main page 7 5 ...
Page 76: ...7 6 ...
Page 77: ...7 6 Calibration Sequence 7 7 ...
Page 78: ...7 7 Calibration Options 7 8 ...
Page 79: ...7 8 Alarm Options 7 9 ...
Page 80: ...7 8 1 Sensor Alarm Options 7 10 ...
Page 81: ...7 9 Instrument Options 7 11 ...
Page 150: ...6 3 Pièces de rechange pour l appareil 6 3 ...
Page 232: ...6 3 Piezas de repuesto 6 3 ...