Sensores IR
Sensor
Rango de
Resolución
Repro-
Repro-
Tiempo de
IR
Pantalla
cibilidad
cibilidad
respuesta a
del punto cero
del valor .
20 °C
medido
t50
t90
CO2
0 - 5% Vol
0.01% Vol
≤ ± 0.05% Vol
≤ ±5%
≤12 s
≤36 s
CO2
0 - 10% Vol
0.01% Vol
≤ ± 0.1 % Vol
≤ ±4%
≤12 s
≤35 s
CO2
0 - 50% Vol
0.1% Vol
≤ ± 1.0 % Vol
≤ ±4%
≤14 s
≤36 s
CH4
0 - 25% Vol
0.2% Vol
≤ ± 0.5 % Vol
≤ ±5%
≤12 s
≤35 s
CH4
0 - 100% Vol
1% Vol
≤ ± 5 % Vol
≤ ±10%
≤11 s
≤34 s
C3H8
0 - 100% LEL
1% LEL
≤ ± 3 % LEL
≤ ±8%
≤11 s
≤32 s
C3H8
0 - 25% Vol
0.1% Vol
≤ ± 0.5 % Vol
≤ ±4%
≤12 s
≤35 s
C3H8
0 - 100% Vol
1% Vol
≤ ± 3 % Vol
≤ ±8%
≤12 s
≤36 s
C4H10
0 - 100% LEL
1% LEL
≤ ± 3 % LEL
≤ ±8%
≤11 s
≤32 s
C4H10
0 - 25 % Vol
0.1% Vol
≤ ± 0.5 % Vol
≤ ±4%
≤12 s
≤35 s
C4H10
0 - 100 % Vol 1% Vol
≤ ± 3 % Vol
≤ ±8%
≤12 s
≤36 s
SENSORES TOXICOS ADICIONALES
Sensor
Rango
Resolución
SO2
0-20 ppm SO2
0.1 ppm SO2
NO2
0-20 ppm NO2
0.1 ppm NO2
NH3
0-100 ppm NH3
1 ppm NH3
PH3
0 - 5.0 ppm PH3
0.05 ppm PH3
HCN
0 - 30.0 ppm HCN
0.5 ppm HCN
Cl2
0-20.0 ppm Cl2
0.1 ppm Cl2
ClO2
0-1.00 ppm ClO2
0.02 ppm ClO2
5-4
Summary of Contents for altair 5
Page 68: ...6 3 Instrument Replacement Parts 6 3 ...
Page 71: ...7 Appendix Flow Charts 7 1 Start Up Sequence Power ON 7 1 ...
Page 72: ...7 2 Normal Operation 7 2 ...
Page 73: ...7 3 Options from Main Page 7 3 ...
Page 74: ...7 4 Password Protected Options 7 4 ...
Page 75: ...7 5 Menu Sequence from Main page 7 5 ...
Page 76: ...7 6 ...
Page 77: ...7 6 Calibration Sequence 7 7 ...
Page 78: ...7 7 Calibration Options 7 8 ...
Page 79: ...7 8 Alarm Options 7 9 ...
Page 80: ...7 8 1 Sensor Alarm Options 7 10 ...
Page 81: ...7 9 Instrument Options 7 11 ...
Page 150: ...6 3 Pièces de rechange pour l appareil 6 3 ...
Page 232: ...6 3 Piezas de repuesto 6 3 ...