• At the end of the FAS, the instrument displays
"FRESH AIR SETUP PASS" or
"FRESH AIR SETUP FAIL".
If the FAS fails, perform a zero calibration per Section 3.8.1.
3.3. Measurement Mode (Normal Operation)
The following Options pages can be executed from the Normal
Operation screen (for further information see Section 2.4 and 7.5):
Peak Page*
shows the peak readings for all sensors
Min Page
shows the minimum reading for the oxygen sensor
STEL Page*
shows the calculated STEL readings of the instrument
TWA Page*
shows the calculated TWA readings of the instrument
Date Page
shows the date
Last Cal
shows the date of the last calibration. If the instrument does not
Date
have a valid calibration, it will display "LAST CAL INVALID"
Cal Due*
shows the set date for the next calibration
Motion Alert
TM
+
if the Motion Alert Feature is activated
* The display of these pages can be de-activated through MSA Link (per Section 2.4)
3.4 Instrument Setup
The instrument allows the user to access and modify the following
parameters through direct button interface:
• Calibration Options
• Alarm Options
• Instrument Options.
These menus can be accessed only from the Measure page by
pressing and holding the ▲ and ▼ buttons simultaneously until
prompted for a password.
The operation is as follows:
1. Turn the instrument ON and wait until the Measure page appears.
2. Press the ▲ and ▼ buttons simultaneously.
• The default password is "672".
3-5
Summary of Contents for altair 5
Page 68: ...6 3 Instrument Replacement Parts 6 3 ...
Page 71: ...7 Appendix Flow Charts 7 1 Start Up Sequence Power ON 7 1 ...
Page 72: ...7 2 Normal Operation 7 2 ...
Page 73: ...7 3 Options from Main Page 7 3 ...
Page 74: ...7 4 Password Protected Options 7 4 ...
Page 75: ...7 5 Menu Sequence from Main page 7 5 ...
Page 76: ...7 6 ...
Page 77: ...7 6 Calibration Sequence 7 7 ...
Page 78: ...7 7 Calibration Options 7 8 ...
Page 79: ...7 8 Alarm Options 7 9 ...
Page 80: ...7 8 1 Sensor Alarm Options 7 10 ...
Page 81: ...7 9 Instrument Options 7 11 ...
Page 150: ...6 3 Pièces de rechange pour l appareil 6 3 ...
Page 232: ...6 3 Piezas de repuesto 6 3 ...