Following a successful calibration of >100% LEL IR sensors, the
unit must be turned OFF, then ON to clear the over-range alarm
"XXX" on the combustible LEL sensor.
3.8.4 Calibration Failure
If the instrument cannot calibrate one or more sensor(s), it goes into the
Calibration Failure page and remains in alarm for 10 seconds.
Sensors that could not be calibrated are indicated by dashed lines on
the concentration display.
If a sensor repeatedly fails calibration after the full calibration procedure
in this manual has been performed, replace the failed sensor.
3.8.5 Calibration with GALAXY Test System
The instrument can also be calibrated using the GALAXY Automated
Test System - contact MSA for a list of compatible gases and
concentrations. The instrument will return a
on the display when all
sensors are successfully calibrated.
3.9 Instrument Shutdown
For instrument shutdown press and hold the button.
• The instrument:
• displays a blinking "HOLD BUTTON FOR SHUTDOWN"
• a progress bar showing the user how much longer to hold the
button to complete the shutdown.
Figure 3-9. Instrument Shutdown
3-20
Summary of Contents for altair 5
Page 68: ...6 3 Instrument Replacement Parts 6 3 ...
Page 71: ...7 Appendix Flow Charts 7 1 Start Up Sequence Power ON 7 1 ...
Page 72: ...7 2 Normal Operation 7 2 ...
Page 73: ...7 3 Options from Main Page 7 3 ...
Page 74: ...7 4 Password Protected Options 7 4 ...
Page 75: ...7 5 Menu Sequence from Main page 7 5 ...
Page 76: ...7 6 ...
Page 77: ...7 6 Calibration Sequence 7 7 ...
Page 78: ...7 7 Calibration Options 7 8 ...
Page 79: ...7 8 Alarm Options 7 9 ...
Page 80: ...7 8 1 Sensor Alarm Options 7 10 ...
Page 81: ...7 9 Instrument Options 7 11 ...
Page 150: ...6 3 Pièces de rechange pour l appareil 6 3 ...
Page 232: ...6 3 Piezas de repuesto 6 3 ...