14.2 Calibration for 5100-26-IT
NOTE: Supplement to Instructions in Section
.
1. The calibration of the Model # 5100-26-IT HF gas sensor module is accomplished by using a 10 ppm Chlorine
standard (Type C kit with # 1260-06 cylinder). Scale range is 0-10 ppm with 5 ppm being mid-scale. 10 ppm Cl2 is
equivalent to 10 ppm HF.
2. Reference Section
for the calibration procedure sub menu. This procedure requires the
use of the magnetic wand to activate the keys and confirm each entry step. Follow prompts on the module display as
navigation is detailed in Section
.
3. Enter actual span gas concentration that is printed on the cylinder being used when prompted.
4. Refer to the sensor rebuild and electrolyte replacement procedure as detailed in Section
.
66
SMC 5100-XX-IT Toxic Gas Detector Module