Function
Display
Description
Reference
Down
*Sentry
Low Alarm Relay set to Sentry
Gas Range Adjustment Example
Enter
Range
S.P. Function - Range Adjust
*C
Enter
*100 PPM
Select [E] to select or ▲ or ▼ to select another and press [E]
Down
10 PPM
Press [E] if selecting 0-10 PPM range
Down
USER
Press [E] if selecting User adjustable range
Enter
100 PPM
Select [E] to select or ▲ or ▼ to select another and press [E]
4-20 mA Adjustment Example
Enter
Calib
S.P. Function - Calib Adjust
*D
Enter
Out: 4mA
Use ▲ or ▼ keys to select 4 mA or 20mA
Enter
4mA
Selects 4 mA
Use ▲ or ▼ keys to adjust the value of the output current (in either 4mA or
20mA menu)
Enter
ACK
Momentary Acknowledge of new Set Point
Mode
Calib
S.P. Function - Calib Adjust
Down
CalibOut
S.P. Function - Output during Calibration Adjust
Enter
Track
Use ▲ or ▼ keys to select Track, C2.50mA, C4.00mA
Track = Output during calibration tracks the calibration gas, C2.50mA = Output during
calibration is 2.50 mA, C4.00mA = Output during calibration is 4.0 mA
Enter
*Track
* = Current selection
Mode
Calib
S.P. Function - Calib Adjust
Down
Tbl Out
S.P. Function - output during Trouble Alarm
Enter
*T1.50mA
Current Value 1.5mA for Trouble
Down
User mA
S.P. Function - User Selectable mA Value
Enter
*T1.50mA
Use ▲ or ▼ keys to select mA Value between 0.5 and 3.7mA
Enter
*T1.50mA
Select 1.5 mA value during trouble
RS-485 Adjustment Example
Enter
Address
S.P. Function - RS-485 Address Adjust
*E
Enter
Addr:016
Use ▲ or ▼ keys to enter new address
Enter
ACK
New address selected
Enter
Address
S.P. Function - RS-485 Address Adjust
Down
Baud
S.P. Function - RS-485 Baud Rate Adjust
Enter
*38400
Press [E] to select or ▲ or ▼ to select another
NOTE: The Key is the same as the function. Mode=M, Enter=E, Up=▲, and Down=▼.
SMC 5100-XX-IT Toxic Gas Detector Module
29
US
5 Operation
Summary of Contents for 5100-03-IT
Page 5: ...15 5100 88 IT Dimensions 67 SMC 5100 XX IT Toxic Gas Detector Module 5 US...
Page 20: ...mA Circuit Types 20 SMC 5100 XX IT Toxic Gas Detector Module 4 Installation US...
Page 50: ...12 Modbus Memory Map 50 SMC 5100 XX IT Toxic Gas Detector Module 12 Modbus Memory Map US...
Page 51: ...13 HART SMC 5100 XX IT Toxic Gas Detector Module 51 US 13 HART...
Page 52: ...52 SMC 5100 XX IT Toxic Gas Detector Module 13 HART US...
Page 68: ......