6.4
Sensor Exposure to Gas
Calibration gas must be delivered to the sensor using the flow rate and duration listed below:
Model
Calibration Gas
Flow
Period
5100-03-IT Air
300 cc of Zero Air or exposure to Ambient Air (3 minutes)
5100-04-IT Carbon Monoxide 300 cc/min
Until Stable (minimum 3 minutes)
5100-05-IT Hydrogen Sulfide 300 cc/min
Until Stable (minimum 3 minutes)
5100-06-IT Chlorine
300 cc/min
Until Stable (minimum 3 minutes)
5100-08-IT* Chlorine Dioxide
300 cc/min
Until Stable (minimum 3 minutes)
5100-10-IT Sulfur Dioxide
300 cc/min
Until Stable (minimum 3 minutes)
5100-12-IT Nitrogen Dioxide
300 cc/min
Until Stable (minimum 3 minutes)
5100-25-IT Ammonia
300 cc/min
Until Stable (minimum 3 minutes)
5100-26-IT Chlorine
300 cc/min
Until Stable (minimum 3 minutes)
5100-88-IT Carbon Dioxide
300 cc/min
Until Stable (minimum 3 minutes)
5100-04-IT Carbon Monoxide 300 cc/min
Until Stable (minimum 3 minutes)
* Use 5PPM Cl
2
to calibrate (set module to 1.5 PPM ClO
2
)
6.5
Calibration Sub-Menu
NOTE: The Key is the same as the function. Mode=M, Enter=E, Up=▲, and Down=▼.
Function
Display
Description
Reference
XXXPPM
Default Display
Mode
ALMRSET:
Mode Function - Alarm Reset
Mode
CALIB:--
Mode Function - Calibrate
Enter
CAL-0PPM-- Banner: Apply zero gas, enter <E> when done (not applicable for 5100-03-IT)
*
Operation: Confirm area clear of gas, or apply zero air to sensor
Enter
ACK
Zero gas setting acknowledged
10PPM-SP
Banner: Select span, enter <E> to calibrate sensor
Banner: (5100-03-IT) Specify Cal gas <E> when done
Sub A
Enter
C 4 PPM
Banner: Apply span gas, then enter <E> to calibrate gas sensor
Sub B
C 4 PPM
Operation: Apply calibration gas
CXXXPPM
Operation: As gas is applied the reading will increase - wait 3 minutes
Enter
CAL-OK
Calibration Passed - now remove gas
WAIT-300
Operation: Five minute time out before sensor is returned to service.
Sub-Routines
10PPM-SP
Banner: Select span, enter <E> when done
Sub A
Down
25PPM-SP
Operation: Change Span Gas Value to 25PPM
Down
Adj-SPAN
Operation: User adjustable value
Enter
25PPM-SP
Operation: Ready for user adjustment
Enter
C 25PPM
Banner: Apply 25PPM gas, then enter <E> to calibrate gas sensor
Sub B
SMC 5100-XX-IT Toxic Gas Detector Module
33
US
6 Calibration
Summary of Contents for 5100-03-IT
Page 5: ...15 5100 88 IT Dimensions 67 SMC 5100 XX IT Toxic Gas Detector Module 5 US...
Page 20: ...mA Circuit Types 20 SMC 5100 XX IT Toxic Gas Detector Module 4 Installation US...
Page 50: ...12 Modbus Memory Map 50 SMC 5100 XX IT Toxic Gas Detector Module 12 Modbus Memory Map US...
Page 51: ...13 HART SMC 5100 XX IT Toxic Gas Detector Module 51 US 13 HART...
Page 52: ...52 SMC 5100 XX IT Toxic Gas Detector Module 13 HART US...
Page 68: ......