5.4
Set-Up
The sensor module set-points menu is used to initially set-up the alarm set points, relay actions, gas type and range, 4-20
mA action or RS-485/Sentry address and baud rates, Digital Input and Warm-up time. Use the [▲] or [▼] keys to select
following menus and press [E] to select.
•
Alarms – Use the [▲] or [▼] keys to select Low Alarm (Warning) or High Alarm (Alarm) menu. Key [▲] will adjust the
setpoint upwards and Key [▼] will adjust the value downwards. Once it reaches the desired setpoint, Key [E] will accept
it and ACK will appear.
Set-points can be configured using this menu up to the following values:
Factory Alarm Set Points
Model
Warning
Alarm
5100-03-IT
19.5%
16.5%
5100-04-IT 50 PPM
100 PPM
5100-05-IT 10 PPM
20 PPM
5100-06-IT 2.0 PPM
5.0 PPM
5100-08-IT
0.3 PPM
1.0 PPM
5100-10-IT 10 PPM
20 PPM
5100-12-IT 2 PPM
4 PPM
5100-25-IT
25 PPM
50 PPM
5100-26-IT
2.0 PPM
5.0 PPM
5100-88-IT
1000 PPM 2500 PPM
NOTE: To enable rising O2 Alarm set warning setpoint higher than 20.9%
•
Relays – Use the [▲] or [▼] keys to select High Alarm (Alarm) or Low Alarm (Warning) relay menu and press [E]. Use
the [▲] or [▼] keys to select the correct alarm relay action for the application, Latch, Sentry or Non-Latch. Selecting
“Sentry” enables the Sentry controller to make all alarm action decisions. * indicates the current selection.
•
Range – Use the [▲] or [▼] keys to select Range menu and press [E]. When “Range” is selected menu provides any
choices of ranges available for the gas type selected. Use the [▲] or [▼] keys to select the desired range. If the “User”
range is selected, use the [▲] or [▼] keys to adjust the high end of the range desired.
Factory Range Set-up
Model
Range
Other Available
5100-03-IT
5-25%
0-25%
5100-04-IT 0-500 PPM
Variable
5100-05-IT 0-100 PPM
Variable
5100-06-IT 0-10 PPM
None
5100-08-IT
0-3 PPM
None
5100-10-IT 0-100 PPM
Variable
5100-12-IT 0-20 PPM
None
5100-25-IT
0-100 PPM
Variable
5100-26-IT
0-10 PPM
None
5100-88-IT
0-5000 PPM Variable
SMC 5100-XX-IT Toxic Gas Detector Module
27
US
5 Operation
Summary of Contents for 5100-03-IT
Page 5: ...15 5100 88 IT Dimensions 67 SMC 5100 XX IT Toxic Gas Detector Module 5 US...
Page 20: ...mA Circuit Types 20 SMC 5100 XX IT Toxic Gas Detector Module 4 Installation US...
Page 50: ...12 Modbus Memory Map 50 SMC 5100 XX IT Toxic Gas Detector Module 12 Modbus Memory Map US...
Page 51: ...13 HART SMC 5100 XX IT Toxic Gas Detector Module 51 US 13 HART...
Page 52: ...52 SMC 5100 XX IT Toxic Gas Detector Module 13 HART US...
Page 68: ......