The hardware diagnostic options (“HW Diag”) are listed below. Press the [E] key to select any of the sub-menus.
Function
Display
Description
Reference
Unlock
sequence
SW Diag
Software diagnostics sub-menu.
Up
HW Diag
Hardware diagnostics sub-menu.
Enter
TempSnsr
Displays the temperature of sensor assembly in °F.
Up
Test WDT
Tests function of internal watchdog.
Up
A2D Test
Displays the values of the 4 ADC channels on the 1-wire interface
(A/B/C/D).
Up
Sen-Init
Initialize sensor assembly for appropriate electrochemical sensor. Use the
[▲] or [▼] keys to select from a series of pre-defined toxic sensor types.
Up
ClearMem
Clears all internal non-volatile memories and restores factory defaults.
The gas detector module will reboot and power up in uncalibrated state.
Up
Run Time
Displays the time gas detector module has been on power.
Up
A2D
Display internal diagnostic voltages. Use the [▲] or [▼] keys to select the
display of the following channels:
0) 4-20mA input signal
1) Digital input switch
2) Module function
3) PCB revision
4) NOT USED
5) Extender Type (Interface Card Type)
6) 5V supply voltage status
Up
HART ID
Displays the HART Device ID. This menu can be used to change the
HART ID.
SMC 5100-XX-IT Toxic Gas Detector Module
31
US
5 Operation
Summary of Contents for 5100-03-IT
Page 5: ...15 5100 88 IT Dimensions 67 SMC 5100 XX IT Toxic Gas Detector Module 5 US...
Page 20: ...mA Circuit Types 20 SMC 5100 XX IT Toxic Gas Detector Module 4 Installation US...
Page 50: ...12 Modbus Memory Map 50 SMC 5100 XX IT Toxic Gas Detector Module 12 Modbus Memory Map US...
Page 51: ...13 HART SMC 5100 XX IT Toxic Gas Detector Module 51 US 13 HART...
Page 52: ...52 SMC 5100 XX IT Toxic Gas Detector Module 13 HART US...
Page 68: ......