Microplate
Reader
Operation
Manual
Chapter 2 Features
─
2
─
Chapter 2
Features
Working conditions
:
Ambient temperature: 10
C~40
C
The relative humidity: 30%~80%(No condensation)
Power: AC100-240V
50-60Hz
2A
Parameters:
Model
Parameter
UT-200A
Light source
Xenon flash lamp >10
9
flashes
Wavelength
200
~
1000nm
Bandwidth
≤ 2.5nm
Wavelength accuracy
2nm
Wavelength
repeatability
0.2nm
Read-out range
0.0-4.0 OD
Linearity@450nm
R2 ≥ 0.999 , [0.0 - 3.0]
Accuracy@450nm
±(1.0% + 0.003A) , (0 - 2.0]
±2.0%
, (2.0 - 2.5]
Precision@450nm
CV < 0.5% (Precision mode); CV < 1.0% (Fast mode)
Stability@450nm
< 0.005A , (0.0 - 2.0]
< 2%
, (2.0 - 2.5]
Measurement speed
<
8 seconds at Fast mode (96-well plate)
<
28 seconds at Precision mode (96-well plate)
Plate shaking
Linear
Incubation range
RT+ 4°C to 45°C
Temp. uniformity
± 0.5°C @ 37°C
Connections
1 B-type of USB port for PC
1 Ethernet port
2 A-type of USB ports for devices
Power requirements
DC24V 6.67A 160W
Dimension (W
×
D
×
H)
300
×
500
×
260mm
Weight (kg)
15.5kg