Chapter 4 Installation
─
4
─
Chapter
4
Installation
Opening
check
.1
Each
UT
-200
A
is
thoroughly
tested
before
shipping,
but
please
check
again
when
you
:receive
the
instrument
and
contact
your
local
distributor
or
manufacturer
if
• The outer package is damaged
• The outer package has any obvious moisture stains
• The outer package has marks of impact
• The outer package has signs of being opened
After opening, please check the instrument and box contents.
Confirm that all ordered accessories have been included.
Check the instrument’s appearance for any damage.
2. Installation
Working condition: locate instrument on a flat dry and clean work table, keep the front
side with enough space for plate holder in and out, also keeping 15cm space for back, left
and right side to enable put or connect wires.
Working environment:
a. Clean air free from corrosion steam or smoke.
b. Temperature should be within the range of +10°C ~ +40°C.
c. Relative humidity should be within the range of 30% ~ 80% to avoid condensation.
Note: KEEP INSTRUMENT AWAY FROM DESTRUCTIVE GAS OR LIQUID!
3. Installation steps
1
Place the instrument on a stable and level surface.
Note: Please DO NOT loose any screw or parts without permission, or it will
cause instrument damage and make it out of warranty.
2
Connect the instrument to an appropriate power outlet using the provided power
cord.
Fig. 1