Microplate
Reader
Operation
Manual
Chapter 5 Operation Guide
─
16
─
5. Result processing
The interface will stay in “Run” interface after detecting samples in “Run” interface and
displays the original absorbance measured under the current protocol. If users want results
been analyzed, just switch to “Result” interface in the side bar, see Fig 13.
Fig 13
Result displayed different according to different layout of protocol settings and detection
modes, the above Fig 13 is results of endpoint. Here, according to different types of
detection modes, the Result is divided into three modes : Endpoint, Kinetic and Spectrum.
5.1 Endpoint result
As Fig 13 shows, data processing type of endpoint result includes “Raw Data”, “Blank
Subtraction”, “Basic Calculation”, “Standard Curve”, “Classification” and “Quality Control”.
Raw data: displays absorbance values of each well, users can switch different
wavelengths by
pressing the button “
” in the middle upper of the main
interface.
Blank Subtraction: blank absorbance is obtained according to blank samples, then
subtract the blank absorbance for each well.
Note:
1. Protocol layout interface must have blank well, otherwise the button is
not available.
2. If a protocol is set with blank sample, the absorbance values of “Basic
Calculation”, “ Standard curve”, “Classification”, “Quality Control” and
“Kinetic Analysis” are all values subtracted blank.