Microplate
Reader
Operation
Manual
Chapter
6
Maintenance,
storage,
transportation
─
24
─
Chapter 6
Maintenance, storage, transportation
1. Maintenance
Keep storage environment dry and clean to prevent moisture, corrosion, away
from strong electromagnetic interference sources.
Instrument already calibrated before leave factory. User is not allowed to
disassembly and make adjustment. Any defectiveness, please contact
manufacturer.
Continuous emergency turning on/off is not allowed.
Make sure apply the device with correct input voltage scope.
Maintenance list
Content
/Day
/Wee
k
/Year
When
needed
Make sure the instrument power off correctly
√
Keep the instrument away from dust
√
Remove overflowing solution right away in case
any damage, then clean it by deionized-distilled
water.
√
If the surface been contaminated with a
biohazard, sterilize it by mild disinfectant.
√
Clean instrument enclosure regularly.
√
Clean the plate holder when necessary.
√
Verification by using light absorption verification
plate.
√
Sterilize the instrument when re-installing or
maintaining.
√
Maintenance
√
2. Storage and transportation
Storage at room temperature -10
℃~
40
℃
, relative humidity less than 80%,
without corrosive gas and with good ventilation.
Keep away from heavy shock, vibration, and humidity during transportation.