Hydraulic Connection
51
0.001
0.01
0.1
1
10
100
1
10
100
1000
Differential
Pressure
(Bar)
Ac
ross
Ports
Viscosity (cSt)
400ml/minute (max. flow)
200ml/minute (ideal flow)
20ml/minute (min. flow)
Figure 15.4
Differential Pressure vs Fluid Viscosity, for various flow rates
15.2 Manual Flow Control
Another possibility is to fit a simple manual flow control (flow restrictor) to the
outlet of the ICM.
•
This should only be done where the available pressure is less than twice the
maximum value calculated. This is because the small orifice size needed to
control the flow from a pressure larger than this has a risk of blockage.
•
The flow controller must be fitted to the outlet only. If fitted to the inlet it will
have a filtering effect.
•
The flow controller must be fitted directly to the ICM outlet port.
15.3 Active Flow Control
This is only needed for operation at high differential pressures, where a too-high
flow would otherwise be generated.