36
Settings
is greater than the lower limit. If the result is in between, both outputs are turned
on and the LED colour will be amber (i.e. a mixture of red and green light).
Alarm Mode 3: Particles-Water
Output 1
Output 2
Turns on When
Cleanliness>Upper Water>Upper
Turns off when
Cleanliness≤lower Water≤lower
Intended Function Cleanliness Alarm Water Alarm
This is used when separate alarm outputs are needed for particles (cleanliness) and
water content.
Alarm Mode 4: Continue - Clean
Output 1
Output 2
Turns on When
Testing
Cleanup Complete
Intended Function Continue Testing Cleanup complete
This is used for a "cleaning’’ application where a signal is needed to run a pump
(``Continue Testing’’) and optionally operate an indicator (``Cleaning Complete’’).
``Continuous testing’’ and ``Stop Testing When Clean’’ must both be selected for
this mode to operate.
Output 1 then operates immediately so it can be used to switch on a pump. Testing
continues until a clean result is obtained. Testing then stops. Output 1 is turned off
and Output 2 turned on to indicate ``Cleaning complete’’
Alarm Mode 5: Tested-not Clean
Output 1
Output 2
This mode is able to use both upper and lower limits such that the outputs have ``hysteresis’’. If this is
not desired then
both upper and lower limits should be set to the same value.