38
Installation
13 Installation
Each ICM supplied consists of the following:
•
ICM
•
Calibration certificate
•
LPA-View CD ROM, software package
•
Pre-wired cable
Optional Equipment:
•
Circular connector pre-wired with 3m cable
•
ICM-RDU Remote display unit
•
500 µm coarse screen filter
•
ICM-FC1 Flow Control Valve
•
ICM-USBi USB adaptor with pre-wired ICM cable
13.1 Installation Procedure
•
Decide on tapping points in hydraulic circuit.
•
Locate the unit mechanically and bolt to desired location using fixing holes pro-
vided. The ICM must be in a vertical orientation, with the oil flowing upwards
through it.
•
Wire back to junction box.
•
Check flow in acceptable range. There needs to be a differential pressure placed
across the ICM, such that a flow of fluid is generated within the range of the
unit.
•
If there is no suitable differential pressure available, then a flow controller will
be needed. One solution is the ICM-FC1 pressure compensated flow control
valve. This limits the flow to around 200ml/minute for differential pressures up
to 400 bar. This should be fitted to the drain side of the ICM (the top fitting).
•
Fix mechanically.
Summary of Contents for ICM