Issue 03 – June 2018
51
If you do not have an anemometer, then you will need to judge the position of speed control knob as
follows:
Knob position
Approx. Air Volume
(m³/hour)
0
0
¼ Turn
275
½ Turn
550
¾ Turn
825
Full Turn (Maximum)
1100
Maintenance
Changing the pre-filters
Regular checks should be carried out to ensure the airflow performance is maintained. Should the
module fail to achieve adequate performance, this normally indicates that the Pre-Filter/s and / or
HEPA filter require changing.
(For replacement filter part numbers, see page 5)
1.
Turn the system off and disconnect from mains.
2.
Remove the screws securing the pre-filter retaining frame.
3.
Remove the pre-filter frame and filter.
4.
Replace pre-filter with the correct type (see page 5).
5.
Replace the screws securing the pre-filter retaining frame.
6.
Reconnect the mains supply and switch on.
7.
Check the airflow and adjust if necessary (as detailed in the setup section)
Changing the HEPA filter
1.
Turn the system off and disconnect from mains.
2.
Remove the screws securing the filter retaining frame.
3.
Remove the filter frame, cover filter (if fitted) and HEPA filter.
4.
Replace HEPA filter with the correct type (see page 5).
5.
Replace cover filter and filter frame using the screws.
6.
Reconnect the mains supply and switch on.
7.
Check the airflow and adjust if necessary (as detailed in the setup section)
8.
Filter should be DOP tested to verify filter integrity and results recorded.
CAM1000 Replacement Filter Part Numbers
Replacement filters are available from Monmouth Scientific:
Pre-Filter
Part No-
PF-0081
HEPA – filter
Part No-
HF-0127
Summary of Contents for Cleanroom Kit 4M
Page 1: ...Issue 03 June 2018 1 Cleanroom Kits ...
Page 2: ...Issue 03 June 2018 2 Installation Guide ...
Page 3: ...Issue 03 June 2018 3 ...
Page 24: ...Issue 03 June 2018 24 Standard Wall Panel ...
Page 25: ...Issue 03 June 2018 25 Door Panels ...
Page 27: ...Issue 03 June 2018 27 Ceiling Assembly 4M 2m x 2m room ...
Page 28: ...Issue 03 June 2018 28 4Mplus 2m x 2m room with 1m x 2m atrium ...
Page 29: ...Issue 03 June 2018 29 6M 3m x 2m room ...
Page 30: ...Issue 03 June 2018 30 6Mplus 3m x 2m room with 1m x 2m atrium ...
Page 35: ...Issue 03 June 2018 35 Using CAM1000 in 9M 9M 12M 12M ...
Page 36: ...Issue 03 June 2018 36 Top Door Runner Assembly ...
Page 40: ...Issue 03 June 2018 40 5 ...
Page 58: ...Issue 03 June 2018 58 6M 3m x 2m room ...
Page 59: ...Issue 03 June 2018 59 6Mplus 3m x 2m room with 1m x 2m atrium ...
Page 60: ...Issue 03 June 2018 60 9M 3m x 3m room ...
Page 61: ...Issue 03 June 2018 61 9Mplus 3m x 3m room with 1m x 3m atrium ...
Page 62: ...Issue 03 June 2018 62 12M 4m x 3m room ...
Page 63: ...Issue 03 June 2018 63 12Mplus 4m x 3m room with 1m x 3m atrium ...
Page 65: ...Issue 03 June 2018 65 6M 3m x 2m room ...
Page 66: ...Issue 03 June 2018 66 6Mplus 3m x 2m room with 1m x 2m atrium ...
Page 67: ...Issue 03 June 2018 67 9M 3m x 3m room ...
Page 68: ...Issue 03 June 2018 68 9Mplus 3m x 3m room with 1m x 3m atrium ...
Page 69: ...Issue 03 June 2018 69 12M 4m x 3m room ...
Page 70: ...Issue 03 June 2018 70 12Mplus 4m x 3m room with 1m x 3m atrium ...