General Information
Chapter Five: Type R7xxC Relay System
52
Trip Point Parameters
Trip Point Values
The trip point values are adjusted using the Trip Point Adjustment pots located on the side of the
Relay Module (refer to Figure 13). The adjustment voltage ranges from 0 to 10 V, corresponding
to zero to full scale. The pots are initially configured to 5.5 V; that is, 55% of FS.
Figure 13: Location of the Trip Point Adjustment Pots
15-Pin Male Type "D"
Analog Output Connector
Trip Point A
Status LED
Trip Point B
Status LED
B
B
Trip Point A
Adjustment Pot
Trip Point B
Adjustment Pot
Trip Point
Test Jacks
Cover Seam