Setup
Chapter Two: Installation
32
Electrical Information for a Voltage Unit
The 7xxC-series transducers require an external power source capable of supplying the voltages
listed in
Appendix A: Product Specifications,
page 65. Noise and ripple should be less than 2
mV (peak-to-peak) over a 10 kHz bandwidth. You may use any readout device capable of
reading from -0.6 V to 11 V. Refer to Figure 9, page 32, for the power, signal and chassis
grounding scheme for a voltage unit.
Note
The ground of any external power supply and readout should be the same
as the transducer ground (chassis ground) to minimize any possible
ground loops and power supply noise which can affect the performance
and stability of the system.
Figure 9: Electrical Scheme for a Voltage Unit
Readout
Power
Supply
+
-
0 to 5 V or 0 to 10 V Pressure Output
Pressure Return
+10.8 to +32 V (for 0 to 5 V)
Power Return
10K
Note 2:
The absolute value of the potential difference between the chassis ground and
signal common/power ground should not exceed 14 Volts.
Note 1:
For best results, use a readout with fully differential inputs.
Multiple Ground
Tie-Point
0.3 F
+13 to +32 V (for 0 to 10 V)
or
Chassis
Ground
Sensor Port
Summary of Contents for Baratron 740C
Page 6: ......
Page 8: ......
Page 10: ......
Page 30: ...Customer Support Chapter One General Information 20 This page intentionally left blank...
Page 55: ...Chapter Two Installation How To Check the Transducer Zero 45...
Page 58: ......
Page 60: ...General Chapter Four Operation 50 This page intentionally left blank...
Page 82: ...General Description Appendix A Product Specifications 72 This page intentionally left blank...