Chapter Two: Installation
Setup
31
Note
If the unit is considerably out of calibration, the zero and span adjustments
could interact slightly. It may take more than one calibration cycle to
accurately complete the adjustments. Always recheck the zero calibration
after adjusting the span. Refer to
Span Adjustment
, page 44, for more
information.
Figure 8: Electrical Scheme for the 4 to 20 mA Unit
Voltage
Readout
Power
Supply
+
-
Return
Excitation
Chassis
Ground
Sensor
Port
13 to 36 V
+
-
-
+
R
sampling resistor
LN