Setup
Chapter Two: Installation
28
Fittings
The 7xxC Series transducers are available with a variety of different fittings, shown in
Table 7:
Fittings Available*
7xxC Single-Ended Configuration
7xxC Flow-Through Configuration
½” Tube
¼” Tube
Swagelok
4-VCR
Male and Female
Swagelok 4-VCR Fixed Male
Swagelok 8-VCR Female
Swagelok 4-VCR Female
NW16-KF **
* Other configurations are available on request. Minimum quantities may be required.
** Up to 1200 Torr only
Table 7: Fittings Available
Caution
1. MKS
does not warranty
the 7xxC Series transducers when single
or double metal ferrule compression-type vacuum fittings are
used because damage will occur to the transducer when improper
tightening procedures are followed.
2. Before proceeding to the
Setup
section, page 23, carefully check all
plumbing connections to the instrument to ensure a
leak-tight installation.
NW16 -KF Fitting Information
The NW16-KF fittings are only available for units with full scale pressure of a maximum of 1200
Torr (23 psia).
Warning
Units with NW-16-KF fittings and a full scale range greater
than 1200 Torr (23 psia or 5 psig) require an HPS overpressure
ring. Operating the unit without a protective overpressure ring
may result in injury.
The HPS part number for the overpressure ring is HPS 100316301.
Summary of Contents for Baratron 740C
Page 6: ......
Page 8: ......
Page 10: ......
Page 30: ...Customer Support Chapter One General Information 20 This page intentionally left blank...
Page 55: ...Chapter Two Installation How To Check the Transducer Zero 45...
Page 58: ......
Page 60: ...General Chapter Four Operation 50 This page intentionally left blank...
Page 82: ...General Description Appendix A Product Specifications 72 This page intentionally left blank...