ELECTRICAL AND DATA CONNECTIONS
Section I: External Input and Output (Ext. I/O)
Connectors
Connector for Unit:
D-Sub 25 Pins, Male
Connector for User Device:
D-Sub 25 Pins, Female M2.6
Recommended Parts:
Connector
HDBB-25S (Hirose Electric Co., Ltd.)
Case
HDB-CTH (Hirose Electric Co., Ltd.)
NOTE:
Be sure to use a connecting cable with a shield line and terminate the shield line appropriately.
Input Signals
Pin No.
Signal
Signal description
1
+24 V output
Power supply for external I/O biasing.
2 Marking
start
*l
When input is closed, marking will start.
3
Marking stop
When input is closed, all marking will stop.
4
Laser start (Laser enable)
When input is closed, all systems are enabled and activated during start-up.
5
Guide beam
When input is closed, the guide beam turns
ON
.
6
Reset errors (Trouble reset)
When input is closed, error (trouble) condition is reset.
7
Stop serial counter
*2
When input is closed the serial counter stops counting (either up or down).
9
External Emergency Stop
When input is opened, all marking operations are stopped.
10
DI common
Common Input terminal for the input signals.
19
0 V
Ground for +24 V DC output.
24 Counter
reset
*3
When input is closed, the serial counter is reset to either the minimum or
maximum value.
25
Remote/Local Control
Switching
When input is closed, the Laser switches to ‘Remote’ (I/O) control.
8, 20, 23
Reserved
Do not use
*1 This signal is valid when the mode is set to “Remote”. Signal is invalid when mode is set to “Local”.
*2 Stops all serial counters.
*3 The signals for the individual counters can be enabled or disabled.